Requirements For Reading
- LAS1000 image
- 1. Sample preparation
- 1) Sample: Human IgG tagged with HRP
- 2) Reagent: ECL+
- 3) Conditions: Immediately after the reaction
- 2. Detection conditions
- 1) Lens aperture: f=0.85
- 2) Exposure time: 10min
- 3) Sample tray position: 3
- 4) Dark frame subtraction: YES
- 5) Flat frame correction: No
- 6) Binning: No
- 3. Image contrast adjustment
- 1) Image type: Negative
- 2) Range: 0-2199
|